Medical Imaging

G20

  • Commodity Name: Diagnostic Color Doppler Ultrasound

 

  • Model: G20

 

  • Construction: This equipment is mainly composed of a main unit, probe and monitor. Optional probes: Convex probe 35C50JA, linear probe 75L40JA, micro-convex probe 35C20GA,65C15JA, transvaginal probe 65C10JA.
  • ApplicationScope:This instrument is applicable to diagnosis of organs such as liver, gall-bladder, kidney, spleen, pancreas, heart, thyroid gland, breast, uterus, bladder, ovary, etc. It has been widely used in examining abdomen and obstetrics, urology, cardiology, gynecology, small parts etc.
  • ManufactureDate: See product packaging for

 

  • ServiceLife:6years (mainunit),2years (probe).

ProbeWorking

Frequency

2.0MHz~12.0MHz

DisplayMode

B,2B,4B,M,B+M,CFM,PDI,DPDI,B+PW,B+CFM+PW,B+PDI+PW

TypeofFocusing

Dynamicfocusing,acousticlensfocusing,launchingmultipointfocusing

ScanningTechnique

Dynamicapodization,dynamicaperture,dynamicfrequencyscanning,

multi-acousticbeam

Pre-processing

lownoisepreamplifier,TGC,filtering,frameaverage,lineaverage

 

Post-processing

Gammacorrection,histogram,digitalscanconversion(DSC),edge enhancement,noiserejection,smooth,ePure,greyscaletransformation,

pseudo-color,colorpersistenceetc

 

DisplayControl

Freeze/Unfreeze,Left-rightreverse,Up-downreverse,Polarityreverse, picturerotation(90°/180°/270°),colorreverse,inverse-frequency spectrum,

pseudo-color.

Control of Transformationand Adjustment on

SoundField

 

output acoustic power, PRF, focus position, scanning angle, image frame frequency, impulse duration, depth, sampling area dimension

GrayScale

256

ColorScale

24

FrameRate

Max.upto70f/s

ScanningArea

≥320mm

DensityofScanning

Lines

Max256line/frame

Monitor

12.1"LCDmonitor

VideoOutput

CVBS,S-video,VGA

DigitalScan

Converter

628´440´24bits

BodyMark

123bodymarkswithprobelocation

 

 

 

 

General Measurement

2D/CFM:Distance,area,volume(ellipsemethod),angle;

M:Distance,time,slope,heartrate,simplifiedleftventriclefunction, complete left ventricle function;

PW: Distance, max ventricular gradient, average ventricular gradient, time, S/Dration,bloodflowrate,accelerationofbloodflow,heartrate,

pulsate&dragindex.

ImageMemory

Harddiskformassiveimagestorage,min.10,000imagespermanentstored

Cineloop

nolessthan1024frames

Biopsy

Optionalbiopsyguide

Character

Date,time,patientname,devicename,usernameetc,user-defined

annotationtable,arrowheadandbodymarks

Measurement Software

Software    packages    of    Abdomen,    Urology,   Cardiology,                  Obstetrics, Gynecology,SuperficialParts,BloodVessels,FetalHeart,andOrthopedic

Surgery;directlyformdiagnosticreportbasedonmeasurementresults.

InputPower

80VA

ContinuousWorking

Time

>8h

Size

395mm(L)X170mm(W)X350mm(H)

Weight

N.W.:≤6.3 kg

 

                TechnicalParametersofStandardConfiguredProbe

 

 

ProbeModel No.

35C50JA

ProbeFrequency

2.0—5.0MHz

ProbeType

Electronicconvex

This product has no descriptive video!